Abstract We introduce a novel evaluation approach for smart bed systems that continuously measure vital signs. In particular, we demonstrate that estimation accuracy (or error) and measurement coverage time are key performance metrics, describing a performance tradeoff in practical smart bed systems. Based on a typical smart bed system that uses a force sensor array placed between bed mattress and frame, we evaluate the effect of different signal filtering options to illustrate viable design choices using our accuracy-coverage tradeoff analysis. In a full night recording study with six participants focusing on respiration rate estimation, we show that measurement coverage is an essential metric that should be analysed together with accuracy, when assessing the performance of smart bed systems.
Date 2014
Proceedings Title Adjunct Proceedings of the 2014 ACM Conference on Pervasive and Ubiquitous Computing (SmartHealthSys '14)
Publisher ACM
Pages 1249–1256
Series Ubicomp Workshop
DOI 10.1145/2638728.2638810
Extra Workshop on Smart Health Systems and Applications